MRI Imaging (Survey, Setting the image capture range,

Return to Summary > Imaging Manual 
Subject Enrollment

Open “Patients” tab,

choose “New Examination

Enter subject information

Patients name:ID identified your own experiment_ subject number.
Registration ID:same as above
(ex: 2001_A_01)

The same ID should be used throughout the experiment, and depending on the experimental design, case-specific symbols should be used so that you can be distinguished in hindsight.

Survey

Click “Start Scan

to start Survey(Imaging for head positioning)

When the following error message appears, press “OK”.

  The patient position on the table is unknown. 
  Please use lightvisor.

Take out the bed once, align the bed again, and then perform the Survey again.

Double click the yellow mark on the left of Examcard of next session.  

Select imaging area

Move the yellow frame to the area to be imaged.
The size is not adjustable as it is subject to ExamCard requirements.
The angle is adjustable.

When you have decided,

Click “Accept

Click “Start Scan

For functional images, a confirmation screen is displayed.

Click “OK

to start imaging.

Examcard turns blue when imaging is completed.

If noise is visible in Auto View, reconnect the head coil connector.
(Unplug it and plug it in again).
Check Auto View again (you can see it while imaging),
if the noise is still noticeable, call Philips.